Commissioning, operation – WIKA PSD-31 User Manual
Page 26
26
WIKA operating instructions pressure switch model PSD-3x
GB
11430486.06 02/2012 GB/D/F/E
Service data (SPDU service protocol data unit)
Service data is always acyclic and exchanged on the request of the IO-Link Master
With the help of the service data, the following parameter values or instrument status can be read:
IO-Link-specific
6. Commissioning, operation
Manufacturer-specific
Index
(decimal)
Object name Format
Access Factory setting
Remarks
16
Vendor name
Visible string
R
WIKA Alexander Wiegand
SE & Co. KG
17
Vendor text
Visible string
R
www.wika.com
19
Product ID
Visible string
R
e.g. 1013093
This SPDU determines the relation-
ship with the corresponding IODD.
21
Serial number
Visible string
R
S#
Corresponds to serial number on
the product label (S#).
24
TAG
max. 16 Byte
Visible string
R/W
-
Customer-specific measuring point
number
33
Last event
Visible string
R
-
40
Process data In 16 Bit unsigned R
-
Display of the process data that are
cyclically transferred.
243
Order no
Visible string
R
P#
Corresponds to product number on
the product label (P#).
Index
(decimal)
Object
name
Format
Access Value range
Factory setting
Remarks
65
SP1 or FH1 16 Bit
unsigned
R/W
1,040 … 9,000
9,000
Switch point/window
High switching output 1
66
RP1 or FL1 16 Bit
unsigned
R/W
1,000 … 8,960
8,200
Reset point/window Low
switching output 1