Stirrer Shaft and Agitator Assembly
The stirrer shaft and agitators are assembled and installed when shipped. Use the following instructions should you need
to remove or adjust these components.
Components Needed for Stirrer Shaft and Agitator Assembly:
1ea CG-2097
Stirrer Shaft
2ea CG-2095-50 Upper PTFE Agitators
1ea CG-2095-55 Lower PTFE Agitator with Stabilizing Bushing
1. The lower agitator assembly (CG-2095-55) is placed on the end of the stirrer shaft. The sand blasted portion is the
lower end of the stirrer shaft.
2. The upper agitator assemblies (CG-2095-50) slide over the end of the shaft. The stirrer shaft has indents that are
spaced every 10 inches from the bottom for the CG-2095-55 agitator, for positioning. Set screws on the agitator
must be aligned with the indents on the stirrer shaft.
CG-2095-50
Upper Agitator,
2ea Supplied
CG-2095-55
Lower Agitator,
1ea Supplied
CG-2097
Stirrer Shaft
CG-2095 Lower
Agitator has Bushing to
Accept a Stabilizing Pin
from Drain Valve