Users Manual for DryBath Block Heaters Models: CLS-3959-01, CLS-3959-02 and CLS-3959-04

Chemglass’s Digital Dry Baths are highly accurate, microprocessor controlled, dry block heating units capable of heating one, two or

four aluminum blocks or aluminum or glass pans. They provide unsurpassed temperature uniformity and accuracy for heating

microtubes, test tubes and other small vessels, including microplates and slides. Each dry bath has a bright green LED display, easy

to set temperature and time controls and comes complete with a block lifter.

SPECIFICATIONS:

CLS-3959-01

CLS-3959-02

CLS-3959-04

Well Size:

1 Block

2 Blocks

4 Blocks

Temperature Range:

Ambient +5 to 150°C

Ambient +5 to 150°C

Ambient +5 to 130°C

Increments and Display Resolution: 0.1°C

0.1°C

0.1°C

Temperature Uniformity:

+/-0.2°C

+/-0.2°C

+/-0.2°C

Temperature Accuracy:

+/-0.2°C

+/0.2°C

+/-0.2°C

Timer Function/Resolution:

Off (untimed) or On to 999 Minutes in 1 Minute Increments

Size W x D x H (in)/Weight:

7.9 x 9.1 x 3.2 (2.2kg)

8.7 x 10.7 x 3.2 (2.5kg)

8.7 x 14.3 x 3.2 (3.5kg)

Electrical 115/230VAC 50/60Hz:

100 Watts

150 Watts

300 Watts

CAUTION: Hot surfaces, especially on the block, can cause serious injury or burns.

CAUTION: Do not put water or liquids into the well as shock, serious injury and death may occur.

CAUTION: Do not heat flammable or explosive substances as serious injury and death may occur.

SET UP and OPERATION

Place the dry bath on a stable, flat surface away from air vents and any equipment exhaust vents. Plug the dry bath into a properly

grounded outlet of the correct voltage. Insert the block(s) into the heating well using the block lifter.

Caution: blocks may be hot!

Using the switch on the back of the unit, turn the power on. The dry bath will make a “beep” sound, illuminate each digit on the

display left to right and briefly illuminate the LED lamps while it performs a unit test. After the unit test, the display will begin showing

the block temperature and the unit will begin heating the block to the previously set temperature and flashing the

Heating lamp.

Setting the Temperature

Use the

Up and Down arrow buttons to increase or decrease the temperature setting. The temperature can be set in tenths of a

degree C. Once the temperature is set, the display will then revert to showing the block temperature and the

Heating lamp will flash

until the set temperature has been reached. The temperature setting is automatically remembered if power is turned off or if the

power is lost.

Using and Setting the Timer

The dry bath has a built-in, independent, digital timer function that sounds a “beeping” alarm when the set time has been reached.

The timer is completely independent of the heating function and does

NOT turn off the heater when set time is reached. This allows

the Timer to alert the user that the time setting has been reached but does not disturb samples being heated. (continued on next