3c. specification chart, D. warranty/service the myron l p, Model p – Myron L PS9TK User Manual
Page 7: Model ps9

3
C. Specification Chart
Parameters
Ranges
Resolution
accuracy
auto
Temperature
compensation
adjustable
Temperature
compensation
cond/TDS
Ratios Pre-
programmed
Conductivity
0-9999µS/cm
10-200mS/cm
in 5 autoranges
0.01(<100µS)
0.1(<1000µS)
1.0(<10mS)
0.01(<100mS)
0.1(<200mS)
±1% of
reading
0-71ºC
32-160ºF
0-9.99%/ºC
KCl, NaCl,
442™
TDS
0-9999ppm
10-200ppt
in 5 autoranges
0.01(<100ppm)
0.1(<1000ppm)
1.0(<10ppt)
0.01(<100ppt)
0.1(<200ppt)
±1% of
reading
0-71ºC
32-160ºF
0-9.99%/ºC
MIN/SALT*
0-9999ppm
10-200ppt
in 5 autoranges
0.01(<100ppm)
0.1(<1000ppm)
1.0(<10ppt)
0.01(<100ppt)
0.1(<200ppt)
±1% of
reading
0-71ºC
32-160ºF
0-9.99%/ºC
pH
0-14pH
±.01pH
±.01pH
**
0-71ºC
32-160ºF
ORP
±999mV
±1mV
±1mV
**
Free
Chlorine
(FC
E
)
0.00-9.99ppm***
0.01ppm
±0.3ppm
<1.00ppm
**
±0.2ppm
≥1.00ppm
**
0-71ºC
32-160ºF
350≤ORPmV<725
and 0.0≤pH<9.9
725≤ORPmV<825
and 0.0≤pH<8.9
Alkalinity Titration
10-800ppm
0.1(<100 ppm)
1(<800 ppm)
Hardness Titration
0-1710ppm
(0-100 grains)
0.1(<100 ppm)
1(<1710 ppm)
LSI Titration
-10 to +10
0.1
Temperature
0-71ºC
32-160ºF
0.1ºC/F
±0.1ºC
* NaCl — Sodium chloride **±.2 pH in the presence of RF fields ≥ 3 V/m and > 300 MHz
***If either ORP or pH is outside the specified limits, the instrument will display "-Or-".
D. Warranty/Service
The Myron L P
OOL
P
RO
, excluding the pH/ORP sensor, has a Two (2)
Year Limited Warranty. The pH/ORP sensor has a Six (6) Month Limited
Warranty for materials and workmanship. If an instrument fails to operate
properly, see Troubleshooting Chart, pgs. 50-51. The battery, pH/ORP
sensor and cell extender are user-replaceable. For other service, return
the instrument prepaid to the Myron L Company.
MyROn L cOMPany
2450 impala Drive
carlsbad, ca 92010-7226 uSa
+1-760-438-2021
E-Mail: [email protected]
www.myronl.com
If, in the opinion of the factory, failure was due to materials or workmanship,
repair or replacement will be made without charge. A reasonable service
charge will be made for diagnosis or repairs due to normal wear, abuse
or tampering. This warranty is limited to the repair or replacement of the
P
OOL
P
RO
only. The Myron L Company assumes no other responsibility
or liability.
E. P
OOL
P
RO
Model
P
OOL
P
RO
Model
PS9
PARAMETERS
Conductivity/MIN/SALT/TDS/Alkalinity/Hardness/LSI/ORPmV/
Free Chlorine Equivalent (FC
E
) ppm/pH/Temperature