cELL chEcK
The cell check verifies the cleanliness of the conductivity/MIN/
SALT/TDS sensor. In normal use the cell may become dirty or
coated and require cleaning. If the display is showing “.00” when
the cell cup is dry, the sensor is probably clean. No matter what a
manufacturer claims, a sensor can and will become contaminated or
coated and, therefore, require cleaning. A true 4-wire sensor, as in
the P
OOL
P
RO
, helps to mitigate contamination, but NO SENSOR IS
100% IMMUNE.
1.
Press
.
2.
Press
to display the stored memory records.
3.
Press
repeatedly until you pass the FAC SEL location.
The display will show a “cELL ch” (see Figure 18).
4.
Press
to test. If cell is clean, “good” will momentarily be
displayed (see Figure 19). If cell is dirty, “cELL cLn” will be
displayed (see Figure 20) (ref. Cleaning Sensors, pg. 48).
Xiv. auTO OFF
Auto off allows the user to adjust the time the instrument is ON (up to
75 seconds) after each press of a key. Default time is 15 seconds with
60 seconds in “caL” (calibration) mode.