Myron L PS9TK User Manual
Page 46

42
4.
Press
to switch units.
5.
Press
to accept unit preference for all temperature
readings.
nOTE: Tempco will still be shown in %/°C.
Xii. TOTaL RETuRn to FacTORy SETTingS “Fac SEL”
There may come a time when it would be desirable to quickly reset all
the recorded calibration values in the instrument back to the factory
settings. This might be to ensure all calibrations are set to a known
value, or to give the instrument to someone else free of adjustments
or recorded data for a particular application.
nOTE: all stored data will be lost.
1.
Press
.
2.
Press
to display the stored memory records.
3.
Press
repeatedly until you pass the “cL
r
aLL” and the
C-F locations. The display will show a “Fac SEL” (see Figure 17).
4.
Press
to accept the resetting. Display will return to
Conductivity.
Figure 16
Figure 15
Figure 17