Myron L PS9TK User Manual
Page 29

25
C. Calibration of Each Solution Type
There is a separate calibration for each of the 4 solution types. Note
that calibration of a 442 solution does not affect the calibration of a
NaCl solution. For example: Calibration (ref. Conductivity, MIN/SALT
or TDS Calibration, pg. 27) is performed separately for each type
of solution one wishes to measure (ref. Conductivity/TDS Standard
Solutions, pg. 52).
D. Procedure to Select a Solution
nOTE: Check display to see if solution displayed (KCl, NaCl or 442)
is already the type desired. If not:
1.
Press
,
MIN
SALT
or
to select the parameter on
which you wish to change the solution type.
2.
Press and hold
key for 3 seconds to make “SEL”
appear (see Figure 1). For demonstration purposes, all 3
solution types are shown simultaneously.
Figure 1
KCl
442
NaCl
3.
Use the
or
key to select type of solution desired
(ref. Solution Characteristics, pg. 57). The selected solution
type will be displayed: KCl, NaCl or 442.
4.
Press
to accept new solution type.