31

Corrective Action

Check connections or replace battery

(ref. Battery Replacement, pg. 27).

1. Recalibrate instrument.

2. Thoroughly rinse sensor well.

3. Recalibrate using fresh buffers (ref. pH

Buffer Solutions, pg. 32).

Replace pH/ORP sensor (ref. Replacement

pH/ORP Sensor, pg. 33).

Clean and rejuvenate sensor (ref. Cleaning

Sensors, pg. 28) and recalibrate. If no

improvement, replace pH/ORP sensor

(ref. Replacement pH/ORP Sensor, pg. 33).

Clean and rejuvenate sensor (ref. Cleaning

Sensors, pg. 28) and recalibrate. If no

improvement, replace pH/ORP sensor

(ref. Replacement pH/ORP Sensor, pg. 33).

Clean cell cup and electrodes (ref.

Cleaning Sensors, pg. 28).

Clean cell cup and electrodes (ref.

Cleaning Sensors, pg. 28).