beautypg.com

Myron L PS6 User Manual

Page 14

background image

10

B. Measuring pH

1.

Remove protective cap by squeezing its sides and pulling up.

2.

Rinse sensor well 3 times with sample to be measured. Shake

out each sample to remove any residual liquid.

3.

Refill both sensor wells with sample.

4.

Press .

5.

Note value displayed.

6.

IMPORTANT: After use, fill pH/ORP sensor well with Myron L

pH Sensor Storage Solution and replace protective cap. If

Myron L pH Sensor Storage Solution is unavailable, you can

use a strong KCl solution, a pH 4 buffer, or a saturated solution

of table salt and tap water (ref. Cleaning Sensors, 2. pH/ORP,

pg. 28).

Do not allow pH/ORP sensor to dry out.

D. Measuring ORP

1.

Remove protective cap by squeezing its sides and pulling up.

2.

Rinse sensor well 3 times with sample to be measured. Shake

out each sample to remove any residual liquid.

3.

Refill both sensor wells with sample.

4.

Press .

5.

Take reading.

NOTE: When ppm free chlorine units are selected (ref. free chlorine unit

selection, pg. 43), annunciators alert you when the concentration is outside

the specified measurement range. “Or” (over range) will display when the

concentration is over the range limit (> 9.9 ppm). “Ur” (under range) will

display when the concentration is below the range limit (< 0.2 ppm).

6.

IMPORTANT: After use, fill pH/ORP sensor well with Myron L

pH Sensor Storage Solution and replace protective cap. If

Myron L pH Sensor Storage Solution is unavailable, you can

use a strong KCl solution, a pH 4 buffer, or a saturated solution

of table salt and tap water (ref. Cleaning Sensors, 2. pH/ORP,

pg. 28).

Do not allow pH/ORP sensor to dry out.

NOTE: FOR INFORMATION REGARDING ThE ORP mV TO

ppm CONVERSION FEATURE, PLEASE SEE SECTION XXVI.

ADDENDUM, PG. 43.