27

testing, we cannot guarantee compatibility of all applications and formats.

We suggest testing your application and format for compatibility before

relying on it.

Additional features, such as assigning a name to the instrument, setting

time and date and erasing data are available. See U2CI software

installation CD or visit our website for the latest instructions:

http://myronl.com/main/U2CI_Application_DL.htm

4.

Upon completion, click on the “disconnect” icon.

5.

Turn off P

ool

P

ro

PC download mode by selecting any

measurement function. Failure to do so will reduce battery life.

* Windows 2000, 2007, XP & Vista are registered trademarks of Microsoft Corporation.

** Macintosh OS9.2 & OSX are registered trademarks of Apple Computer Inc.

XVI. CARE and MAINTENANCE

P

ool

P

ros

should be rinsed with clean water after use. Solvents should

be avoided. Shock damage from a fall may cause instrument failure.

A. Temperature Extremes

Solutions in excess of 71°C/160°F should not be placed in the cell

cup area; this may cause damage. The pH sensor may fracture if the

P

ool

P

ro

temperature is allowed to go below 0°C/32°F. Care should be

exercised not to exceed rated operating temperature.

Leaving the P

ool

P

ro

in a vehicle or storage shed on a hot day can easily

subject the instrument to over 66°C/150°F. This will void the warranty.

B. Battery Replacement

Dry Instrument ThOROUGhLy. Remove the four (4) bottom screws.

Open instrument carefully. Carefully detach battery from circuit board.

Replace with 9 volt alkaline battery. Replace bottom, ensuring the

sealing gasket is installed in the groove of the top half of case. Re-install

screws, tighten evenly and securely.

NOTE: Because of nonvolatile EEPROM circuitry, all data stored in

memory and all calibration settings are protected even during power

loss or battery replacement. However, loss of time and date may occur if

battery is removed for longer than 3 minutes (180 seconds).

C. pH/ORP Sensor Replacement

Order model RPR. When ordering, be sure to include the model and serial

number of your instrument to ensure receipt of the proper type. Complete

installation instructions are provided with each replacement sensor.