Agilent Technologies G3180B User Manual
Page 3

Installation and Operation Guide
3
In this Guide. . .
This Installation and Operation Guide contains information for installing and 
using an effluent splitter on an Agilent 6890 gas chromatograph (GC). The 
G3180 splitter is intended for use with capillary columns and uses makeup gas 
to maintain adequate flows throughout the system.
1
Introduction
This chapter describes how the splitter works, the GC and software 
requirements of the system and the contents of the installation kit.
2
Hardware Installation
See this chapter for a detailed procedure for installing the splitter hardware 
and connecting the makeup gas supply.
3
Splitter Configurations
The split ratio (how the column effluent divides between the two detectors) is 
governed by two restrictors, which are lengths of deactivated fused silica 
tubing. This chapter presents a set of precalculated “typical” configurations. If 
desired, you can create a custom configuration to meet specific needs. The 
chapter describes a set of software tools, included in the kit, to assist you in 
designing such configurations. Finally, installation of the column and 
restrictors is covered.
4
Operation
This chapter contains a worked-out custom configuration, plus a few special 
topics.