Splitting to an msd – Agilent Technologies G3180B User Manual
Page 28

28
Installation and Operation Guide
3
Splitter Configurations
For constant pressure methods, first find the maximum flow as above. Use the 
GC, ChemStation, Flow Calculator Software or the Method Translation 
Software to find the inlet pressure that gives the maximum flow at the upper 
temperature of the method (make sure the column outlet pressure is set to 
3.8 psig for the calculation).
For example, if a 30 m × 0.32-mm id column is used with
Configuration 1, using
helium carrier and programming to 300 °C, the pressure that gives a flow of 
9.4 mL/min ([5.2 + 5.2] – 1 = 9.4) is 56.3 psig. This is the maximum pressure at 
which the inlet should be set. The inlet should not be set at or below 3.8 psig.
If you decide to use a typical configuration, note the restrictor dimensions 
from 
and proceed to
“Restrictor and Column Installation"
Splitting to an MSD
Note that the maximum column flows for an MSD are quite low. This limit is 
imposed by the rating of the turbo or diffusion pump. Configurations with 
split ratios greater than 1 can be used but peak broadening and/or tailing 
should be expected. They are shown in the configuration tables more as a 
caution than as a recommendation.
In practice, the column flow can be set to within 0.5 mL/min of the total flow 
if necessary. For example, the 1:1 split to an MSD with a performance turbo 
pump running a method programmed to 300 °C should have a column flow of 
no more than 2.3 mL/min ([1.4 + 1.4] – 0.5) at 300 °C.
Split ratios to the MSD greater than 1 are very limited due to these flow 
considerations and should be avoided if possible.