Details, Metal ferrules, Microfluidic plate – Agilent Technologies G3180B User Manual
Page 10: Constant pressure operation

10
Installation and Operation Guide
1
Introduction
The column flow mixes with the makeup flow in the splitter. This mixture then 
flows through lengths of uncoated, deactivated, fused-silica tubing to each 
detector. These tubes act as flow restrictors. While the flow through each 
restrictor changes with oven temperature, the ratio of the two flows at any 
temperature is the same.
Details
The G3180B kit addresses several limitations of previous approaches to 
splitting column effluent between two detectors:
Metal ferrules
The splitter uses metal column ferrules, which eliminate air leakage into the 
sample stream. Unlike polyimide, metal ferrules do not loosen upon thermal 
cycling of the oven. They also do not outgas contaminants or shed particles 
(like graphite) that can result in chromatographic problems.
Microfluidic plate
The splitting hardware is based on microfluidic plate technology. This allows 
very low dead volume connections between the column end and the two 
detector restrictor tubes. The thin metal plate has fast thermal response and is 
mounted solidly on the oven wall for ease of use. The interior plate surfaces 
are deactivated to prevent adsorption by active compounds.
Constant pressure operation
The splitter uses a source of makeup gas supplied by electronic pneumatics 
control (EPC). This maintains the splitter at a known and constant pressure. 
Constant pressure allows easier splitting to vacuum detectors like the MSD. It 
simplifies choice of splitter parameters, allowing all aspects of the 
chromatographic setup to be calculated. Constant pressure makeup allows the 
column to be run in constant flow mode while still maintaining a constant 
split ratio between two detectors of different operating pressures such as the 
FPD and the MSD. Because the EPC pressure can be time programmed, useful 
operations like backflushing unwanted heavy materials from the column and 
changing columns in MSD systems without venting are possible.
