Agilent G3180B Splitter Kit
Installation and Operation Guide

Agilent Technologies

3
Splitter Configurations

Typical Configurations

26

Splitting to an MSD

28

Custom Configurations

29

Restrictor id and length

32

Maximum and minimum flows

33

Column outlet pressure

34

Inlet pressure

34

Restrictor and Column Installation

35

Install the column

35

Connect the splitter

35

Disconnect tubing from the splitter

37

The combination of restrictor diameters and lengths determines how the
column effluent is divided (the split ratio) between the two detectors. There
are two approaches to setting up a splitter method.

Use a typical configuration. A set of eight configurations is discussed

beginning on the next page. They apply to a variety of detector
combinations and split ratios. All of the flows have been calculated.

Create a custom configuration. If the typical configurations do not meet

your needs, you can create one that does. The CD shipped with the splitter
kit provides tools for the necessary calculations.

We suggest examining the typical configurations first, since they cover a wide
variety of splitter applications and require no calculations.