Supported Commands
SDI-12 Commands Supported
The following commands are supported by the EnviroSMART and EasyAG SDI-12 probe interfaces:
Command
Name
Response
a!
Acknowledge
Active
a
aI!
Send
Identification
allccccccccmmmmmmvvvxxxxxxxxxxxx
aAb!
Change Address
a
aM!
Start
Measurement
atttn
values (n) up to 9
aMn!
Start
Measurement
n in range “1”
- “3”
atttn
values (n) up to 9
Note: A response of a0000
provided indicating all values are
served by previous Start Measurement
commands (aMn!)
aC!
Concurrent
Measurement
atttnn
to 16
aCn!
Concurrent
Measurement
n in range “1”
- “9”
atttnn
to 16
Note: A response of a0000
provided indicating all values are
served by previous Start Concurrent
Measurement (aCn!) commands
aDn!
Send Data
n in range “0”
- “9”
Refer to “Data Reading” section
aV!
Start
verification
a0000
aRn!
Continuous
Measurement
a
Probe interface will respond with its
address