CS511-L Dissolved Oxygen Probe

2. Cautionary Statements

• The CS511 is a precision instrument. Please handle it with care.

• Because the CS511 is shipped dry, electrolyte needs to be added before

using the probe (see Section 4.1, Getting Probe Ready to Use).

• Letting the CS511 dry up shortens the life of the membrane and probe.

• Drain the solution from the CS511 before storing it out of water.

• Replace the membrane and recalibrate the probe before redeploying the

CS511 after it has been stored out of water or dried up in the field.

3. Initial Inspection

• Upon receipt of the CS511, inspect the packaging and contents for

damage. File damage claims with the shipping company.

• Immediately check package contents against the shipping documentation

(see Section 3.1, Shipping Kit and Accessories). Contact Campbell
Scientific about any discrepancies.

• The model number and cable length are printed on a label at the

connection end of the cable. Check this information against the shipping
documents to ensure the expected product and cable length are received.

3.1 Shipping Kit and Accessories

Except for the agitator, these items are for Sensorex’s
DO6400/T. Refer to Appendix B if you have a DO6200/T.

NOTE

3.1.1 Shipping Kit

(1) Membrane replacement tool
(1) Bottled DO electrolyte, 250 ml
(2) Teflon membranes
(2) Membrane O-rings
(2) Membrane spaces

3.1.2 Optional Probe Accessories

• PT4─L Agitator with Repeat Cycle Timer for stagnant conditions (see

Appendix C)

• 22261 Maintenance Kit containing (5) Teflon membranes, (5) membrane

O-rings, (5) tensioning washers, and a 250-ml bottle of electrolyte

• 22262 Maintenance Kit containing (25) Teflon membranes, (25) membrane

O-rings, (25) tensioning washers, and a 500-ml bottle of electrolyte

• 22263 Spare Parts Kit containing (2) membrane locks, (2) tensioning

washers, (2) body O-rings, and (1) membrane-replacement tool

2