beautypg.com

1 example cr1000 program using voltdiff, Example cr1000 program using voltdiff, 2. wiring for example 1 – Campbell Scientific CS511 Dissolved Oxygen Probe User Manual

Page 18

background image

CS511-L Dissolved Oxygen Probe

7.2.1.1 Example CR1000 Program Using VoltDiff

This example is a CR1000 program but programming for the CR800, CR850,
CR3000, and CR5000 is similar. TABLE 7-2 shows the wiring for the
example.

TABLE 7-2. Wiring for CR1000 Example

CR1000 Connection

Sensor Wire

1H White

1L Black

Ground Clear

'CR1000

'Declare Variables and Units
Public Batt_Volt
Public DOmV
Public DOppm

Units Batt_Volt=Volts
Units DOmV=mV
Units DOppm=ppm

'Define Data Tables
DataTable(Table1,True,-1)
DataInterval(0,60,Min,10)
Sample(1,DOmV,FP2)
Sample(1,DOppm,FP2)
Sample(1,Batt_Volt,FP2)
EndTable

DataTable(Table2,True,-1)
DataInterval(0,1440,Min,10)
Minimum(1,Batt_Volt,FP2,False,False)
EndTable

'Main Program
BeginProg
Scan(5,Sec,1,0)

'Default Datalogger Battery Voltage measurement Batt_Volt:

Battery(Batt_Volt)

'CS511 Dissolved Oxygen Probe measurements DOmV and DOppm:

VoltDiff(DOmV,1,mV250,1,True,0,_60Hz,1,0)
DOppm=DOmV*0.34

'Call Data Tables and Store Data

CallTable(Table1)
CallTable(Table2)
NextScan
EndProg

12