CS511-L Dissolved Oxygen Probe

7.2.1.2 Example CR200(X) Program

The CR200(X)-series must use the VoltSE() instruction since these dataloggers
do not make differential measurements. If the other CRBasic dataloggers use
the VoltSE() instruction instead of the VoltDiff() instruction, their
programming will be similar to this example. TABLE 7-3 shows the wiring for
the example.

TABLE 7-3. Wiring for CR200(X) Example

CR200(X) Connection

Sensor Wire

SE1 White

Ground Black

Ground Clear

'CR200(X) Series

'Declare Variables and Units
Public Batt_Volt
Public DOmV
Public DOppm

Units Batt_Volt=Volts
Units DOmV=mV
Units DOppm=ppm

'Define Data Tables
DataTable(Table1,True,-1)
DataInterval(0,60,Min)
Sample(1,DOmV)
EndTable

DataTable(Table2,True,-1)
DataInterval(0,1440,Min)
Minimum(1,Batt_Volt,False,False)
EndTable

'Main Program
BeginProg
Scan(10,Sec)

'Default Datalogger Battery Voltage measurement Batt_Volt:

Battery(Batt_Volt)

'CS511 Dissolved Oxygen Probe measurements DOmV and DOppm:

VoltSE(DOmV,1,1,1,0)
DOppm=DOmV*0.34

'Call Data Tables and Store Data

CallTable(Table1)
CallTable(Table2)
NextScan
EndProg