CS511-L Dissolved Oxygen Probe

3. Select CS511 Dissolved Oxygen Probe, and select the right arrow (in

center of screen) to add it to the list of sensors to be measured, and then
select Next.

4. Define the name of the public variables and enter the calibration

multiplier. Variables default to DOmv for the millivolt measurements and
DOppm for the ppm values. The default calibration value of 0.34 is based
on an average. It is preferable to calibrate the probe using the procedure
provided in Section 7.3, Calibration. After entering the information, click
on OK, and then select Next.