Appendix C. PT4-L Agitator

C-4

The CR510, CR10X, and CR23X use instruction P86 and P22 as follows:

CR10X P86 and P22 Instructions Example

45: Do (P86)
1: 41

Set

Port

1

High


46: Excitation with Delay (P22)
1: 1

Ex

Channel

2: 20

Delay W/Ex (units = 0.01 sec)

3: 0

Delay After Ex (units = 0.01 sec)

4: 0

mV

Excitation


47: Do (P86)
1: 51

Set

Port

1

Low


48: End (P95)

The above examples are not as power efficient as possible and would require
AC power to maintain a sufficient battery charge. If it is necessary to operate
an agitator without AC power available, write the program so that the agitator
is only operated for a short period of time just before the measurement is to be
taken.