CS511-L Dissolved Oxygen Probe

membrane in a screw-on membrane cap. Regular servicing is not required.
When necessary, the probe can be fully overhauled in five min.

The CS511’s cable can terminate in:

• Pigtails that connect directly to a Campbell Scientific datalogger

(option –PT).

• Connector that attaches to a prewired enclosure (option –PW). Refer

to

www.campbellsci.com/prewired-enclosures

for more information.

• Connector that attaches to a CWS900 Wireless Sensor Interface

(option –CWS). The CWS900 allows the probe to be used in a
wireless sensor network. Refer to

www.campbellsci.com/cws900

for

more information.

6. Specifications

Features:

• In-line thermistor provides automatic temperature compensation

• Agitator available that keeps the probe clean and moves water across

membrane for more accurate readings

• Compatible with all Campbell Scientific dataloggers

Compatible Dataloggers:

CR200(X)-series

CR800

series

CR1000

CR3000

CR5000

CR9000X

CR510

CR10(X)

CR23X

CR7

21X

Principle of Measurement:

Membrane-covered, galvanic oxygen probe

Output Signal:

33 mV

± 9 mV (100% saturation),

< 2 mV (0% saturation)

Accuracy:

Better than

± 2% of reading ± 1 digit when

calibration temperature equals measuring
temperature

± 5°C

Response Time:

5 min. from 100% to 0% oxygen

Materials of Construction:
Body:
Anode:
Cathode:


Noryl
Silver
Zinc

Diameter:

5.72 cm (2.25 in)

Height:

17.78 cm (7 in) from bottom of sensor to end
of cable-strain relief