CS500 Temperature and Relative Humidity Probe

10

Example 1. Sample CR10(X) Program that Computes Vapor Pressure
and Saturation Vapor Pressure

;Measure the CS500 temperature.
;
01: Volt (SE) (P1)

1:

1

Reps

2:

5

2500 mV Slow Range

;CR500 (2500 mV); CR23X (1000 mV); 21X,
CR7 (5000 mV)

3:

3

SE Channel

;Black wire (SE 3), Green wire (G)

4:

1

Loc [ T_C ]

5:

.1

Mult

;See Table 2 for alternate multipliers

6:

-40

Offset

;See Table 2 for alternate offsets

;Measure the CS500 relative humidity.
;
02: Volt (SE) (P1)

1:

1

Reps

2:

5

2500 mV Slow Range

;CR500 (2500 mV); CR23X (1000 mV); 21X,
CR7 (5000 mV)

3:

4

SE Channel

;Brown wire (SE 4), Green wire (G)

4:

2

Loc [ RH_frac ]

5:

.001

Mult

;See Table 3 for alternate multipliers

6:

0

Offset

;Limit the maximum value of relative humidity
;to 1 (expressed as a fraction).
;
03: If (X<=>F) (P89)

1:

2

X Loc [ RH_frac ]

2:

3

>=

3:

1

F

4:

30

Then Do

04: Z=F (P30)

1:

1

F

2:

0

Exponent of 10

3:

2

Z Loc [ RH_frac ]

05: End (P95)

;Compute the saturation vapor pressure in kPa.
;The temperature must be in degrees Celsius.
;
06: Saturation Vapor Pressure (P56)

1:

1

Temperature Loc [ T_C ]

2:

3

Loc [ e_sat ]