beautypg.com

D.2 sdi-12 transparent mode, D.2.1 cr200(x) series datalogger example, D.2.1 – Campbell Scientific CS451/CS456 Pressure Transducer User Manual

Page 41

background image

Appendix D. SDI-12 Sensor Support

TABLE D-2. SDI-12 Commands

SDI-12 Command

Command Function

Values Returned

aD!

Configured settings

Pressure/Level,

Temperature

aD1!

psig, °C

Pressure, Temperature

aD2!

psig, °F

Pressure, Temperature

aD3!

kPa, °C

Pressure, Temperature

aD4!

kPa, °F

Pressure, Temperature

aD5!

Sensor’s Serial Number

Serial Number

aD6!

Ohms, ohms, °C

R, Rb, Temperature

aD7!

Configured settings

(provides data in less than

0.8 seconds)

Pressure/Level

aD8!

Configured settings

(provides average of data

based on user selected

samples)

Pressure/Level,

Temperature

D.2 SDI-12 Transparent Mode

System operators can manually interrogate and enter settings in probes using

transparent mode. Transparent mode is useful in troubleshooting SDI-12

systems because it allows direct communication with probes. Datalogger

security may need to be unlocked before transparent mode can be activated.

Transparent mode is entered while the PC is in telecommunications with the

datalogger through a terminal emulator program. It is easily accessed through

Campbell Scientific datalogger support software, but is also accessible with

terminal emulator programs such as Windows HyperTerminal. Datalogger

keyboards and displays cannot be used.

The terminal emulator is accessed by navigating to the Datalogger menu in

PC200W, the Tools menu in PC400, or the Datalogger menu in the Connect

screen of LoggerNet.

The following examples show how to use LoggerNet software to enter

transparent mode and change the SDI-12 address of a CS451/CS456 sensor.

The same steps are used to enter transparent mode with PC200W and PC400

software after accessing the terminal emulator as previously described.

D.2.1 CR200(X) Series Datalogger Example

1. Connect a single CS451/CS456 to the CR200(X) (see TABLE 7-1).

2. In the LoggerNet Connect screen navigate to the

Datalogger menu and

select

Terminal Emulator. The terminal emulator window will open. In

the

Select Device menu, located in the lower left-hand side of the window,

select the CR200Series station.

D-3