1 every visit, Troubleshooting, Every visit – Campbell Scientific CS451/CS456 Pressure Transducer User Manual

Page 28: Every two to three years or on a rotating schedule

background image

CS451/CS456 Submersible Pressure Transducer

Periodic evaluation of the desiccant is vital for keeping the vent tube dry. The

CS451/CS456 ships with the desiccant tube attached. To assess the

effectiveness of the desiccant, use one of the following:

The desiccant in the tube changes color from blue to pink when the

drying power is lost.

The Enclosure Accessory Humidity Indicator Card (pn 28878).

9.1 Every Visit

Collect data.

Visually inspect wiring and physical conditions.

Check indicating desiccant or enclosure humidity indicator; service if

necessary.

Check battery condition (inspect physical appearance and use a keyboard

display, PDA, or laptop to view the battery voltage).

Check all sensor readings; adjust transducer offsets if necessary.

Check recent data.

9.2 Every Two to Three Years or on a Rotating Schedule

Send the CS451/CS456 in for inspection.

10. Troubleshooting

The most common causes for erroneous pressure transducer data include:

poor sensor connections to the datalogger

damaged cables

damaged transducers

moisture in the vent tube

Problem:

Unit will not respond when attempting serial communications.

Suggestion:

Check the power (red is +V and black is ground) and signal (white is SDI-12

data) lines to ensure proper connection to the datalogger. Check the datalogger

program to ensure that the same port the SDI-12 data line is connected to is

specified in the measurement instruction.

18