1 every visit, Troubleshooting, Every visit – Campbell Scientific CS451/CS456 Pressure Transducer User Manual
Page 28: Every two to three years or on a rotating schedule

CS451/CS456 Submersible Pressure Transducer
Periodic evaluation of the desiccant is vital for keeping the vent tube dry. The
CS451/CS456 ships with the desiccant tube attached. To assess the
effectiveness of the desiccant, use one of the following:
•
The desiccant in the tube changes color from blue to pink when the
drying power is lost.
•
The Enclosure Accessory Humidity Indicator Card (pn 28878).
9.1 Every Visit
•
Collect data.
•
Visually inspect wiring and physical conditions.
•
Check indicating desiccant or enclosure humidity indicator; service if
necessary.
•
Check battery condition (inspect physical appearance and use a keyboard
display, PDA, or laptop to view the battery voltage).
•
Check all sensor readings; adjust transducer offsets if necessary.
•
Check recent data.
9.2 Every Two to Three Years or on a Rotating Schedule
•
Send the CS451/CS456 in for inspection.
10. Troubleshooting
The most common causes for erroneous pressure transducer data include:
•
poor sensor connections to the datalogger
•
damaged cables
•
damaged transducers
•
moisture in the vent tube
Problem:
Unit will not respond when attempting serial communications.
Suggestion:
Check the power (red is +V and black is ground) and signal (white is SDI-12
data) lines to ensure proper connection to the datalogger. Check the datalogger
program to ensure that the same port the SDI-12 data line is connected to is
specified in the measurement instruction.
18