For example, if the correct elevation of the water, as measured by a staff gauge
or other measurement device, is 2015.50 feet, and the CS451 provides a
reading of 5.76 psig, then:
5.76 psig • 2.31 ft/psig = 13.3056 ft.
So, the offset is calculated:
2015.50 ft – 13.3056 ft = 2002.1944 ft
This offset can be accounted for in the program instruction of the SDI-12
recorder.
7.3 Wiring
Power down your system before wiring the CS451/CS456. The
shield wire plays an important role in noise emissions and
susceptibility as well as transient protection.
7.3.1 SDI-12 Datalogger Connections
TABLE 7-1. SDI-12 Wiring
Color
CS451/CS456
Function
CR800
CR5000
CR3000
CR1000
CR200(X)
Series
CR510
CR23X
CR10X
Red
+12 Vdc
12 V
Battery+
12 V
Black
Power Ground G
G
G
White
SDI-12 Signal *Control Port C1/SDI-12 Control Port
Blue
GND
GND
G
G
Yellow GND
GND
G
G
Clear
Shield
GND
G
G
*dedicated SDI-12 port on CR5000
7.3.2 RS-232 Connections
TABLE 7-2. RS-232 Wiring
Color
CS451/CS456
Function
Connection
RS-232 9-pin
Red
+12vdc
Power Source
Black
Power Ground
Power Ground
White
RS-232 Tx (Output) Transmit
Pin 2 Rx (Input)
Blue
RS-232 Rx (Input)
Receive
Pin 3 Tx (Output)
Yellow
Digital Ground
Ground
Pin 5 GND
Clear
Shield GND
Ground