1. General

The CS105 analog barometer uses Vaisala’s Barocap

® silicon capacitive

pressure sensor. The Barocap sensor has been designed for accurate and stable
measurement of barometric pressure. The CS105 outputs a linear 0 to 2.5
VDC signal that corresponds to 600 to 1060 mb. It can be operated in a
powerup or continuous mode. In the powerup mode the datalogger switches
12 VDC power to the barometer during the measurement. The datalogger then
powers down the barometer between measurements to conserve power.

2. Specifications

Operating Range

Pressure:

600 mb to 1060 mb

Temperature: -40

°C to +60°C

Humidity: non-condensing

Accuracy

Total Accuracy***

±0.5 mb @ +20°C

±2 mb @ 0°C to +40°C

±4 mb @ -20°C to +45°C

±6 mb @ -40°C to +60°C

Linearity*:

±0.45 mb @ 20°C

Hysteresis*:

±0.05 mb @ 20°C

Repeatability*:

±0.05 mb @ 20°C

Calibration uncertainty**:

±0.15 mb @ 20°C

Long-Term Stability:

±0.1 mb per year

* Defined

as

±2 standard deviation limits of end-point non-linearity,

hysteresis error, or repeatability error

** Defined

as

±2 standard deviation limits of inaccuracy of the working

standard at 1000 mb in comparison to international standards (NIST)

*** Defined as the root sum of the squares (RSS) of end-point non-linearity,

hysteresis error, repeatability error and calibration uncertainty at room
temperature

General

Dimensions: 9.7 cm x 5.9 cm x 2.1 cm (3.8” x 2.3” x 0.8”)

Weight: 110 g (4 oz)

Housing material: anodized aluminum