CS105/CS105MD Barometric Pressure Sensor
07: Real Time (P77)
1: 0110
Day,Hour/Minute
08: Sample (P70)
1: 1
Reps
2: 1*
Loc [ P_mb ]
-Input Locations-
1 P_mb
*
Proper entries will vary with program and datalogger channel, and input location assignments.
** On the 21X use the 5000 mV slow range.
Example 3. CR10(X) Program for Measuring the CS105 Every Execution Interval
;{CR10X}
;
*Table 1 Program
01: 2
Execution
Interval
(seconds)
01: Do (P86)
1: 41*
Set Port 1* High
02: Excitation with Delay (P22)
1: 1
Ex
Channel
2: 0
Delay W/Ex (units = 0.01 sec)
3: 100
Delay After Ex (units = 0.01 sec)
4: 0
mV
Excitation
03: Volts (SE) (P1)
1: 1
Reps
2: 25**
± 2500 mV 60 Hz Rejection Range
3: 7*
SE
Channel
4: 1*
Loc [ P_mb ]
5: .184
Mult
6: 600
Offset
04: Do (P86)
1: 51*
Set Port 1* Low
05: If time is (P92)
1: 0
Minutes (Seconds --) into a
2: 30
Interval (same units as above)
3: 10
Set Output Flag High (Flag 0)
06: Real Time (P77)
1: 0110
Day,Hour/Minute
07: Average (P71)
1: 1
Reps
2: 1*
Loc [ P_mb ]
8