Appendix A. Daily Average

If the CS105 will be measured every minute or faster, and a daily average is
required, follow Example 5. In Example 5 the daily average, maximum, and
minimum pressure is recorded at midnight. A constant is subtracted from the
pressure before the average instruction is executed. This prevents the loss of
significant digits when adding small numbers to large numbers in floating
point arithmetic.

Example 5. Recording a Daily Average, Maximum, and Minimum Pressure

;{CR10X}
;
*Table 1 Program
01: 5

Execution

Interval

(seconds)


01: Do (P86)
1: 41*

Set Port 1* High


02: Excitation with Delay (P22)
1: 1

Ex

Channel

2: 0

Delay W/Ex (units = 0.01 sec)

3: 100

Delay After Ex (units = 0.01 sec)

4: 0

mV

Excitation


03: Volts (SE) (P1)
1: 1

Reps

2: 25**

± 2500 mV 60 Hz Rejection Range

3: 1*

SE

Channel

4: 1*

Loc [ P_mb ]

5: .184

Mult

6: 600

Offset


04: Do (P86)
1: 51*

Set Port 1 Low*


05: Z=X+F (P34)
1: 1*

X Loc [ P_mb ]

2: -800

F

3: 2*

Z Loc [ P_const ]


06: If time is (P92)
1: 0

Minutes (Seconds --) into a

2: 1440

Interval (same units as above)

3: 10

Set Output Flag High (Flag 0)


07: Set Active Storage Area (P80)
1: 3

Input

Storage

Area

2: 3*

Loc [ P_mb_avg ]