Page 13

background image

CS105/CS105MD Barometric Pressure Sensor

-Input Locations-
1 P_mb

*

Proper entries will vary with program and datalogger channel, and input location assignments.

** On the 21X use the 5000 mV slow range.

Example 4. 21X Program for Measuring the CS105 Every Execution Interval Using an
Excitation Channel to Turn the CS105 On and Off.

;{21X}
;
*Table 1 Program
01: 2

Execution

Interval

(seconds)


01: Excite Delay Volt (SE) (P4)
1: 1

Reps

2: 5

SE

Channel

4: 1*

Excite all reps w/Exchan 1

5: 100

Delay (units 0.01 sec)

6: 5000

mV

Excitation

7: 1*

Loc [ P_mb ]

8: .184

Mult

9: 600

Offset


02: If time is (P92)
1: 0

Minutes

into

a

2: 30

Minute

Interval

3: 10

Set Output Flag High


03: Real Time (P77)
1: 0110

Day,Hour/Minute


04: Average (P71)
1: 1

Reps

2: 1*

Loc [ P_mb ]


-Input Locations-
1 P_mb

* Proper entries will vary with program and datalogger channel, and input location assignments.

5. Long Lead Lengths

There is a 0.06 mV/foot voltage drop in the CS105 signal leads. This voltage
drop, in long lead lengths, will raise the barometric reading by approximately
1.1 mb per 100 feet.

For lead lengths greater than 20 feet, use the differential instruction
(Instruction 2) to measure the CS105.