3 cg3 specifications, Installation – Campbell Scientific CNR1 Net Radiometer User Manual

Page 7

background image

CNR1 Net Radiometer

Tilt response:

Max. dev. 2%

Overall ISO classification:

second class

Sensitivity:

10 - 35 µV/(W m

-2

)

Impedance:

125 Ohm nominal

Operating temperature:

-40°C to +80°C

Spectral range:

305-2800 nm (50% points)

Expected signal range for
atmospheric application:


0 - 15 mV typical

Expected accuracy for daily sums:

± 10%

Window heating offset:

Max. 25 W m

-2

at 1000 W m

-2

normal

incidence solar radiation

2.3 CG3 Specifications

Response time 95%:

18 s

Non-stability:

< 1% change per year

Non-linearity:

Max. dev. 2.5% (0-1000 W m

-2

)

Temperature dependence of
sensitivity:


6% (-10°C to +40°C)

Tilt response:

Max. 3% when facing downwards

Field of view:

150 degrees

Sensitivity:

5 - 35 µV/(W m

-2

)

Impedance:

125 Ohm nominal

Operating temperature:

-40°C to +80°C

Temperature range for specified
behavior:


-10°C to +40°C

Measurement range:

-250 to +250 W m

-2

Spectral range:

5 to 50 µm

Expected signal range for
atmospheric application:


-4 to 4 mV

Expected accuracy for daily sums:

± 10%

3. Installation

For measurement of the Net Radiation, it is most important that the instrument
is located in a place that is representative of the entire region that one wishes to
study.

When installed on a mast, the preferred orientation should be such that no
shadow is cast on the Net Radiometer at any time during the day. In the

3