beautypg.com

Calibration – Campbell Scientific CNR1 Net Radiometer User Manual

Page 26

background image

CNR1 Net Radiometer

;Measure CNR1 temperature

3: Full Bridge w/mv Excit (P9)

1: 1

Reps

2: 22

50 mV, 60 Hz Reject, Slow, Ex Range

3: 22

50 mV, 60 Hz Reject, Slow, Br Range

4: 5

DIFF

Channel

5: 1

Excite

all

reps

w/Exchan

1

6: 4200

mV

Excitation

7: 5

Loc [ Temp_C ]

8: 1

Mult

9: 0

Offset

4: Temperature RTD (P16)

1: 1

Reps

2: 5

R/R0 Loc [ Temp_C ]

3: 5

Loc [ Temp_C ]

4: 1.0

Mult

5: 0

Offset

5: Z=X+F (P34)

1: 5

X Loc [ Temp_C ]

2: 273.15 F

3: 6

Z Loc [ Temp_K ]

;Output data to final storage every 60 minutes

6: If time is (P92)
1: 0

Minutes (Seconds --) into a

2: 60

Interval (same units as above)

3: 10

Set Output Flag High (Flag 0)


7: Real Time (P77)
1: 0220

Day,Hour/Minute

(midnight

=

2400)


8: Average (P71)
1: 6

Reps

2: 1

Loc [ CM3_up ]


Copy range code options from example 1 here:

* Proper entries….

* Proper entries will vary with program and input channel usage.
** 25 mV range for CR10X, 50 mV for 21X and CR7
*** 4200 mV for 21X and CR7, 2100 mV for CR10X
+ 25 mV range for CR10X, 50 mV for 21X and CR7
++ 7.5 mV range for CR10X, 5 mV for 21X and CR7

7. Calibration

The CNR1 should be recalibrated every two years, or as an alternative, by
letting a higher standard run parallel to it over a two-day period and then
comparing the results. For comparison of pyranometers, one should use a clear
day. For comparison of pyrgeometers, one should compare nighttime results.
Deviations of more than 6% can be used to correct the calibration factors.

22