beautypg.com

Campbell Scientific CNR1 Net Radiometer User Manual

Page 14

background image

CNR1 Net Radiometer

The four radiation outputs can be measured using Differential or Single-Ended
inputs on the datalogger. A differential voltage measurement (Instruction 2) is
recommended because it has better noise rejection than a single-ended
measurement. When differential inputs are used, jumper the low side of the
input to AG or

to keep the signal in common mode range.

TABLE 5-1. Datalogger Connections for Differential Measurement, When Using a 4WPB100

Function

Color

CR10X

CR800, CR850,

CR1000

21X/CR7, CR23X

CM3 Up Signal

Red

Differential Input (H)

Differential Input (H)

Differential Input (H)

CM3 Up Reference

*Blue

Differential Input (L)

Differential Input (L)

Differential Input (L)

CM3 Down Signal

White

Differential Input (H)

Differential Input (H)

Differential Input (H)

CM3 Down Reference

*Black

Differential Input (L) Differential

Input

(L) Differential

Input

(L)

CG3 Up Signal

Grey or

†Orange

Differential Input (H)

Differential Input (H)

Differential Input (H)

CG3 Up Reference

*Yellow

Differential Input (L) Differential

Input

(L) Differential

Input

(L)

CG3 Down Signal

Brown

Differential Input (H)

Differential Input (H)

Differential Input (H)

CG3 Down Reference

*Green

Differential Input (L) Differential

Input

(L) Differential

Input

(L)

Shield Shield G

*Jumper to AG or

with user supplied wire.

†Cables extended > 48’ have an orange wire.

Pt-100 Temperature Sensor Connections to 4WPB100 and Datalogger

Function

Color

4WPB100

Datalogger

Black

Wire

Excitation

H

Differential

Input

(H)

Pt-100 Excitation +

Red

L

Differential Input (L)

Pt-100 Excitation -

Blue

G

(AG CR10X/CR510)

Pt-100 Signal +

Yellow

Differential Input (H)

Pt-100 Signal –

Green

Differential Input (L)

TABLE 5-2. Datalogger Connections for Single-Ended Measurement, When Using a 4WPB100

Function

Color

CR10X

CR23X, 21X, CR7

CR800, CR850,

CR1000

CM3 Up Signal

Red

Single-Ended Input

Single-Ended Input

Single-Ended Input

CM3 Up Reference

Blue

AG

CM3 Down Signal

White

Single-Ended Input

Single-Ended Input

Single-Ended Input

CM3 Down Reference

Black

AG

CG3 Up Signal

Grey or Orange

Single-Ended Input

Single-Ended Input

Single-Ended Input

CG3 Up Reference

Yellow

AG

CG3 Down Signal

Brown

Single-Ended Input

Single-Ended Input

Single-Ended Input

CG3 Down Reference

Green

AG

Shield Shield G

10