4 full bridge measurements, Full bridge measurements, 5. four wire half bridge hookup and measurement – Campbell Scientific AM16/32B Relay Multiplexer User Manual

Page 38: 6. full bridge measurement, And section 6.4

background image

AM16/32B Relay Analog Multiplexer

/

0

"4 X 16" Mode

COM H (ODD)

COM L

COM H (EVEN)

COM L

ODD H

ODD L

EVEN H

EVEN L

SHIELD

SENSOR SHIELDS

COM

CR23X/

CR800/
CR850/

CR1000


21X/

CR7



CR10(X)


CR3000/

CR5000

EX E E VX

H H H H

L L L L

AG

H H H H

L L L L

G

FIGURE

The CR5000 and CR3000 also have current excitation channels which allow a

the excitation current is known, it is not

current a
Consider

6-5. Four wire half bridge hookup and measurement

resistance measurement. Because
necessary to measure the voltage across a fixed resistor to determine the

s in FIGURE 6-5. See Section 5.3, General Programming

ations, for an example.

"4 X 16" Mode

COM H (ODD)

COM L

COM H (EVEN)

COM L

ODD H

ODD L

EVEN H

EVEN L

SHIELD

SENSOR SHIELDS

COM

/

0

CR23X/

CR800/
CR850/

CR1000


21X/

CR7



CR10(X)


CR3000/

CR5000

EX E E VX

AG

H H H H

L L L L

6-6. Full bridge measurement

G

FIGURE

6.4 Full Bridge Measurements

Sensor to Multiplexer Wiring — With panel switch set to “4x16” mode,
excitation, ground, and the two signal leads may be connected to one input
SET (FIGURE 6-6).

Multiplexer to Datalogger Wiring — COM terminals are connected to a
datalogger excitation channel, a differential analog input channel, and an

28