Met One Rain Gage Models 380 and 385

3. Pour the appropriate amount of water (see Table 7-1) into the inner funnel

(i.e. 18.52 ml for 0.01 inches of rain / tip):

Conversion Factors:
Tip to in

3

: 113.04

in

2

(catch orifice area) x increment in inches.

in

3

to ml:

multiply by 16.3881.

ml to ounces:

multiply by 0.03382.

4. Turn the cup adjustment screw, opposite the full bucket, up until the

bucket assembly tips. Tighten the lock nut.

5. Repeat steps 3 and 4 for the opposite bucket.

6. Repeat steps 1-5 to ensure proper calibration.

7. Reinstall the housing assembly and tighten the three screws.

TABLE 7-1. Volume of Water

for Recalibration

Tip Increment

in

3

ml

ounces

0.01 in

1.130

18.52 0.63

7.2 Factory Recalibration

Factory recalibration is available from either Campbell Scientific Incorporated
or Met One Instruments. Please call and request an RMA number prior to
returning any equipment to Campbell Scientific Incorporated

15