223 Delmhorst Cylindrical Soil Moisture Block

7.2.1.2 Edlog

The generalized Edlog programming sequence follows:

ACTIVATE MULTIPLEXER/RESET INDEX

For the CR10(X) and CR23X, use Edlog instruction Do (P86) to set
the port high. For the 21X and CR7, use Edlog instruction Set
Port(s) (P20)
to set the port high.

BEGIN MEASUREMENT LOOP

Use Edlog instruction Beginning of Loop (P87)

CLOCK PULSE AND DELAY

With the CR23X and CR10(X) the clock line is connected to a
control port. Instruction Do (P86) with the pulse port command
(71 – 78) pulses the clock line high for 10 ms. Instruction Excitation
with Delay (P22)
can be added following the Do (P86) to delay an
additional 10 ms.

MEASURE SENSOR AND CALCULATE RESISTANCE
See

Section

7.2.2, Excite and Measure the 223, and Section 7.2.3,

Calculate Sensor Resistance.

END MEASUREMENT LOOP

Use Edlog instruction End (P95).

DEACTIVATE MULTIPLEXER

For the CR10(X) and CR23X, use Edlog instruction Do (P86) to set
the port low. For the 21X and CR7, use Edlog instruction Set Port(s)
(P20)
to set the port low.

7.2.2 Excite and Measure the 223

The sensor is excited and measured using the BrHalf instruction in CRBasic or
Instruction 5 (AC Half Bridge) in Edlog. Recommended excitation voltages
and input ranges are given in TABLE 7-2. TABLE 7-2 shows the excitation
and voltage ranges used with our dataloggers.

TABLE 7-2. Excitation and Voltage Ranges

Datalogger

mV Excitation Full Scale Range

CR800/CR850 250

±250 mV

CR1000

250

±250 mV

CR3000

200

±200 mV

CR5000

200

±200 mV

21X

500

±500 mV

CR7

500

±500 mV

CR10(X)

250

±250 mV

CR23X

200

±200 mV

10