Ashcroft 2274 Digital Industrial Gauge User Manual
Page 9
– 9 –
How to Use Your Menu Functions
To set a user password (SETPW):
Step 1: Press the MENU key on the keypad
Step 2: Press the
ᮡ
(up arrow key) or
ᮢ
(down arrow
key) until the word CONFIG appears.
Step 3: Press ENTER. The word SETPW appears on the
gauge display
Step 4: Press ENTER. A five digit numeric password is
now required.
Step 5: Press the
ᮡ
(up arrow key) or
ᮢ
(down arrow
key) on the keypad to select the first digit of the password.
Step 6: Press ENTER.
Step 7: Repeat until the five-digit password is shown on
the gauge display.
Step 8: Press ENTER.
Note: to erase password at any time while in the SETPW
(set password) mode, press the ZERO/CLEAR key. The
user will be prompted to reprogram the password once the
five-digit password is entered. The gauge will display SAVE.
Step 9: Press ENTER to save the password setting.
ENTPW: Once a user password has been established and
entry into the CONFIG mode is required, the user will be
prompted to ENTPW or enter password.
Follow setup steps 4-8 above.
RECAL: or recalibrate allows for zero, mid-scale, full-scale
and factory default calibration of the gauge. The RECAL
feature also allows for recalibration of gauges with
4-20mA output.