INDICATION
LIKELY REASON
REMEDY
calibration off
gas metered is not the same as
use matched calibration
what meter was calibrated for
composition of gas changed
see K factor tables in APPENDIX 2
gas leak
locate and correct
pc board defective
return to factory for replacement
RFE dirty
flush clean or disassemble to
remove impediments
occluded sensor tube
flush clean or disassemble to
remove impediments or return to
factory for replacement
filter screen obstructed
flush clean or disassemble to
at inlet
remove impediments or replace
transducer is not
check for any tilt or change in the
mounted properly
mounting of the transducer;
generally, units are calibrated for
horizontal installation
(relative to the sensor tube)
GFC valve does
incorrect valve adjustment
re-adjust valve (section 7.2.4)
not work
in open position
pc board defect
return to factory for replacement
cable or connectors
check cable and connectors
malfunction
or replace
differential pressure too high
decrease pressure to correct level
insufficient inlet pressure
adjust appropriately
GFC valve does incorrect valve adjustment
re-adjust valve (section 7.2.4)
not work in
closed position
pc board defect
return to factory for replacement
cable or connectors
check cable and connectors
malfunction
or replace
orifice obstructed
disassemble to remove
impediments or return to factory