Aalborg GFC User Manual
Page 24

7.2.14
50% Flow Adjustment
Change the setpoint to 2.50 VDC to control at 50% of full scale flow. Check the
flow rate indicated against the flow calibrator. If the flow rate is not within ±0.75%
of full scale, re-adjust the setting for potentiometer [R38], until the flow output is
correct.
7.2.15
75% Flow Adjustment
Change the setpoint to 3.75 VDC to control at 75% of full scale flow. Check the
flow rate indicated against the flow calibrator. If the flow rate is not within ±0.75%
of full scale, re-adjust the setting for potentiometer [R39], until the flow output is
correct.
7.2.16
100% Flow Adjustment
Change the setpoint to 5.00 VDC to control at 100% of full scale flow. Check the
flow rate indicated against the flow calibrator. If the flow rate is not within ±0.75%
of full scale, re-adjust the setting for potentiometer [R40], until the flow output is
correct.
Repeat steps 7.2.13 to 7.2.16 at least once more.
TABLE II GFC SOLENOID VALVE ORIFICE SELECTION TABLE
7.2.17
LCD Display Scaling
It may be desirable to re-scale the output reading on the LCD readout supplied
with certain model GFC transducers. Re-calibration for a new flow range or dif-
ferent engineering units are two examples of when this may be necessary.
7.2.17.1
Access LCD Display Circuit
Carefully remove the LCD from the GFC or panel mounted surface. Remove the
aluminum housing on the side of the connection cable. Slide the LCD assembly
out of the aluminum housing.
20
ORIFICE PART NUMBER
FLOW RATE [N
2
]
OR.020
10 to 1000 sccm
OR.040
1 to 5 slpm
OR.055
5 to 10 slpm
OR.063
10 to 15 slpm
OR.094
20 to 50 slpm
OR.125
50 to 100 slpm