Jumper Settings
RS-485 serial bus termination can be selected with the on-board jumper.
Termination is on when pins 1 and 2 are shorted, and termination is off
when pins 2 and 3 are shorted or when the jumper plug is fully
disconnected. Termination is implemented by a 120 resistor on the
component board.
In a simple point-to-point configuration where only one WAC155
transmitter is used the jumper is always in ON position. In a multidrop
configuration where several WAC155 transmitters share the single
RS-485 communication line, the termination jumper is in ON position
only at the physically farthest transmitter (i.e. at the end of the
communication line), and in OFF position at the other transmitters.
0701-017
Figure 4
RS-485 Termination