This chapter provides the technical data of the Product.

Technical specifications

Measuring range

GMT221

0 ... 2% CO

2

0 ... 3% CO

2

0 ... 5% CO

2

0 ... 10% CO

2

0 ... 20% CO

2

GMT222

0 ... 2000 ppm
0 ... 3000 ppm
0 ... 5000 ppm
0 ... 7000 ppm
0 ... 10 000 ppm

Accuracy (including repeatability, non-linearity and calibration

uncertainty) at 25 °C and 1013 hPa

GMT221

± (1.5% of range + 2% of reading)

(applies

for

concentrations above 2% of full scale)

GMT222

± (1.5% of range + 2% of reading)

Temperature dependence, typical - 0.3 % of reading / °C (output

decreases when temperature rises,
default 25

C (77 F))

Pressure dependence, typical

+ 0.15 % of reading / hPa (output
increases when pressure rises, default
1013 hPa (1 atm))

Long-term stability

< ± 5 %FS / 2 years

Response time (63%)

GMT221

20

seconds

GMT222

30

seconds

Warm-up time

30 seconds
15 minutes full specification