Technical data, Technical specifications, Chapter 6 – Vaisala GMT220 User Manual
Page 35

Chapter 6 ____________________________________________________________ Technical Data
VAISALA _______________________________________________________________________ 35
CHAPTER 6
TECHNICAL DATA
This chapter provides the technical data of the Product.
Technical specifications
Measuring range
GMT221
0 ... 2% CO
2
0 ... 3% CO
2
0 ... 5% CO
2
0 ... 10% CO
2
0 ... 20% CO
2
GMT222
0 ... 2000 ppm
0 ... 3000 ppm
0 ... 5000 ppm
0 ... 7000 ppm
0 ... 10 000 ppm
Accuracy (including repeatability, non-linearity and calibration
uncertainty) at 25 °C and 1013 hPa
GMT221
± (1.5% of range + 2% of reading)
(applies
for
concentrations above 2% of full scale)
GMT222
± (1.5% of range + 2% of reading)
Temperature dependence, typical - 0.3 % of reading / °C (output
decreases when temperature rises,
default 25
C (77 F))
Pressure dependence, typical
+ 0.15 % of reading / hPa (output
increases when pressure rises, default
1013 hPa (1 atm))
Long-term stability
< ± 5 %FS / 2 years
Response time (63%)
GMT221
20
seconds
GMT222
30
seconds
Warm-up time
30 seconds
15 minutes full specification