Chapter 5 ______________________________________________________________ Maintenance

VAISALA _______________________________________________________________________ 33

10. Select

1-point adjustment

, press

SELECT

. Press

READY

if the value

has stabilized.

11. Give the reference concentration value by using the arrow buttons.

Examples: For a zero-point adjustment using pure nitrogen (N

2

)

enter value 0 ppm.

For other reference gases, the analyzed CO

2

concentration is

typically printed on the bottle. Use that value.

Take care that you give the correct value in correct unit, you
cannot revert the earlier values after accepting the new values.
Press

OK

.

12. Confirm the adjustment, press

YES

. By pressing

NO

you return

to adjustment mode display.

13. Adjustment is complete. Press

BACK-EXIT

to return to the basic

display.

14. Shut off the gas flow.

15. Replace the screw onto the adjusting button.