One-point adjustment procedure, Caution – Vaisala GMT220 User Manual
Page 32

USER'S GUIDE____________________________________________________________________
32 __________________________________________________________________ M010022EN-F
16. Confirm the adjustment, press
YES
. By pressing
NO
you return
to adjustment mode display. (If the difference between the two
references is less than 20% of the measuring range of the probe,
adjustment cannot be done).
17. Adjustment is complete. Press
BACK-EXIT
to return to the basic
display.
18. Shut off the gas flow.
19. Replace the screw onto the adjusting button.
One-Point Adjustment Procedure
CAUTION
Please take special care regarding the following when carrying out the
adjustment:
- Check that you give the correct reference concentrations in the correct
units.
- Check that the reading has really stabilized before accepting the
reading.
You cannot revert to the factory settings after the adjustment!
1.
Insert the probe into the field check adapter (as deep as possible,
the perforated filter cover should be completely inside the adapter).
2.
Connect the tubing to the bottom port of the adapter.
3.
Connect the adapter with the tubing to the flow meter, the pressure
regulator and further to the reference gas bottle. The side port of
the adapter is left open for gas outflow.
4.
Let the reference gas flow and stabilize with a flow rate of 0.6
l/min. Follow the reading of GM70, and after it has stabilized, wait
for 6 more minutes.
5.
Take out the screw from the GM70 probe handle to expose the
adjustment button (see Figure 11 on page 31). Press the button with
a small screwdriver. When pressing the button, the indicator turns
to adjusting mode.
6. Press
OK
to confirm the adjusting.
7. Select
CO2
, press
SELECT.
8. Press
YES
to give the ambient pressure and temperature values.
To continue adjusting press
EXIT
.
9.
Now the adjustment mode is on. Press
GRAPH
to see when the
readings have stabilized. Go back and press
ADJUST
to select the
adjustment method.