Chapter 5 ______________________________________________________________ Maintenance

VAISALA _______________________________________________________________________ 31

0505-236

Figure 11

Location of the GM70 Adjustment Button

6. Press

OK

to confirm the adjusting.

7. Select

CO2

, press

SELECT.

8. Press

YES

to give the ambient pressure and temperature values.

To continue adjusting press

EXIT

.

9.

Now the adjustment mode is on. Press

GRAPH

to confirm that the

readings have stabilized. Go back and press

ADJUST

to select the

adjustment method.

10. Select

2-point adjustment

, press

SELECT

. Press

READY

with the

stabilized reading in the lower reference concentration.

11. Give the lower reference concentration value by using the arrow

buttons (for example, if you are using pure nitrogen, enter value 0
ppm). Take care that you give the correct value in correct units, as
you cannot revert the earlier values after accepting the new values.
Press

OK

.

Next move on to the adjustment at the second (high-end)
reference point.

12. Take out the tubing from the low-end gas bottle and connect it to

the high-end gas bottle.

13. Let the high end reference gas flow and stabilize with a flow rate of

0.6 l/min. Follow the reading of GM70, and after it has stabilized,
wait for 6 more minutes.

14. Press

READY

with the stabilized reading in the higher reference

concentration.

15. Give the high-end reference concentration value by using the arrow

buttons. The analyzed CO

2

concentration of the reference gas is

typically printed on the bottle.
Take care that you give the correct value in correct units, as you
cannot revert to the earlier values after accepting the new values.
Press

OK

.