Error indication in transmitters without display, Error messages – Vaisala GMT220 User Manual
Page 25

Chapter 4 _______________________________________________________Serial Communication
VAISALA _______________________________________________________________________ 25
Error Indication in Transmitters without
Display
Error codes are cabled by blinking a message with short flashes of the
LED lights.
LEDs are situated on the cover of the transmitter. The green
OPERATION
LED
is blinking when a non-critical error occurs (no effect on relays). The
red
LEVEL 1
and
LEVEL 2
LED
s are blinking when a fatal error occurs
(switches the relays OFF).
NOTE
The red level LED is continuously lit when the predefined CO
2
concentration is exceeded.
Error Messages
Table 6
Error Messages
Error
Message
Probable Cause
Action
Severity
Er 01
Main board memory
problem
Contact Vaisala Service
Center
Fatal
Er 02
Main board memory
problem
Contact Vaisala Service
Center
Fatal
Er 03
Main board memory
problem
Contact Vaisala Service
Center
Fatal
Er 04
Probe contact failure Contact Vaisala Service
Center
Fatal
Er 05
Probe contact failure Contact Vaisala Service
Center
Fatal
Er 06
Probe contact failure Contact Vaisala Service
Center
Fatal
Er 08
Incompatible probe
Contact Vaisala Service
Center
Fatal
Er 10
IR-source failure
Contact Vaisala Service
Center
Fatal
Er 11
IR-source failure
Contact Vaisala Service
Center
Fatal
Er 12
Sensor failure
Contact Vaisala Service
Center
Fatal
Er 13
Signal error
None
Non-
critical
Er 14
Measurement out of
range
Check environment and
recalibrate if necessary
Non-
critical