beautypg.com

Merit Medical VP-210 PD Catheter Implantation System User Manual

Page 5

background image

7. Use the radiopaque stripe as a guide to avoid

twisting the catheter. (Figure 24). For optimal catheter

placement, radiopaque stripe should be oriented

directly anterior or directly posterior in the patient.

CAUTION: make sure the catheter is not doubled on itself,

kinked, or twisted.

Figure 24

8. advance catheter until distal cuff reaches rectus sheath.

(There will be an increase in resistance to movement of

the catheter) (Figure 25).

Figure 25

9. Position the Cuff Implantor Tool parallel with and over

the catheter, between the two cuffs (Figure 26).

Figure 26

10. advance Cuff Implantor Tool to edge of distal cuff

(Figure 27).

NOTE: To improve visualization of the cuff, it is helpful to

retract incision site tissue.

Figure 27

11. advance catheter and Cuff Implantor Tool

simultaneously 1.0 cm to both dilate the rectus muscle

and advance the cuff into the rectus muscle while

holding the luke Guide stationary with the hemostat

(Figure 28).

Figure 28

REMOVING TOOLS

1. Verify cuff position visually or digitally.

NOTE: To improve visualization of the cuff, it is helpful to

retract incision site tissue.

2. Retract the luke Guide parallel with the catheter

(Figure 29).

NOTE: maintain pressure on cuff with the Cuff Implantor

Tool to hold it in position.

Figure 29

3. Retract Cuff Implantor Tool, parallel with the catheter,

without dislocating or moving the distal cuff (Figure 30).

Figure 30

4. Retract the catheter stylette.

5. allow any remaining air to exit.

ChECkING CAThETER PATENCY

1. Test catheter patency via infusion of 100-500 cc sterile

saline (Figure 31).

Figure 31

2. If catheter is functioning well, fluid will flow out in a

steady drip or flow when proximal end of the catheter is

lowered below the primary site when the syringe has

been removed.

3. The proximal end of the catheter can also be raised ap

proximately 12-15 cm above the patient’s abdomen.

Fluid will rise and fall within the catheter tube in

conjunction with respiration.

TUNNELING ThE CAThETER

1. locate the previously marked exit-site as determined by

the Implantation Stencil (Figure 32).

Figure 32

This manual is related to the following products: