Figure 4

3. ask patient to tighten abdominal muscles prior to

inserting the luke Guide assembly.

4. Insert luke Guide assembly at a 45° angle from

horizontal, pointing towards the coccyx (Figure 5).

CAUTION: It is important to maintain a 45° angle to assure

proper anchoring in the rectus muscle and final catheter

placement.

Figure 5

ChECkING ThE POSITION

1. Remove trocar from assembly (Figure 6).

Figure 6

2. Insert Y-TEC Peritoneoscope (scope) into cannula and

lock together (Figure 7).

Figure 7

3. Confirm location of distal end of cannula and scope

within the peritoneum (Figure 8).

Figure 8

4. Remove scope (Figure 9).

Figure 9

5. (optional) attach air Insufflation Set, (sold separately, not

available in all areas) as needed to cannula (Figure 10).

Figure 10

6. Place patient in typical Trendelenburg position.

7. Insufflate filtered room air (approximately 700-1200 cc,

depending on patient size).

8. Detach air Insufflation Set. Place thumb or finger on

cannula to retain air.

POSITIONING ThE CAThETER

1. Re-insert peritoneocope (Figure 11).

Figure 11

2. aim distal tip of scope into the air pocket by

making the scope more parallel with the abdomen

(Figure 12).

Figure 12

3. Examine peritoneum to find optimal location for

catheter. note any adhesions or abdominal

characteristics that may impede proper catheter

placement. (Figure 13).

Figure 13