Merit Medical PD Percutaneous Kit IFU User Manual

Page 4

background image

Cautions:

• Verify that the catheter is not twisted or kinked.

• Do not use excessive force as the catheter is pulled.

Excessive force can permanently damage the integrity

of the catheter walls or dislocate the distal part of the

catheter.

• If resistance is noted, carefully use a hemostat inserted

into the primary implantation site to open the tunnel

track for the exit cuff.

• Never use a hemostat at the exit-site.

• DO NOT pull the catheter off the tunneler.

• DO NOT cut the suture off. Attempting to cut the suture

off creates a very high risk of damaging the catheter.

39. Inject a small amount of non-ionic contrast to check

that the catheter placement and assure it is not twisted

or kinked.

40. Close the primary incision with sutures.

41. DO NOT suture the exit-site.

42. Confirm placement by infusing 100 to 1000 mL of

warmed sterile saline.

43. If necessary, use clip included in kit to clamp the

catheter to control the flow of saline.

44. Lock the catheter with 10 mL heparin before applying

connector and cap or transfer set (sold separately, not

available from Merit).

45. According to standard hospital protocol, close the

initial catheter insertion site, attach the appropriate

connector, and wound dressing.

References:

1. Abdel-Aal AK, Fluoroscopic and Sonographic Guidance

to Place Peritoneal Catheters: How We Do It, AJR: 2009;

192:1085-1089

Manufacturer:

Merit Medical Systems, Inc.

1600 West Merit Parkway, South Jordan, Utah 84095 U.S.A.

1-801-253-1600 U.S.A. Customer Service 1-800-356-3748

Authorized Representative:

Merit Medical Ireland Ltd,

Parkmore Business Park West, Galway, Ireland

www.merit.com

ID 030414 401059001/A