Procedures for set-up – AirSep MN152-1 User Manual
Page 8

4
Procedures For Set-Up
Using One SureFlow Oxygen Flow Station
The following step-by-step instructions apply to SureFlow flow stations used with
oxygen concentrators, although you may use any medical-grade oxygen source
that does not exceed 50 psig (345 kPa).
To supply oxygen to the SureFlow unit:
During set-up, nothing is initially connected to the oxygen concentrator at this
time. This is important to prevent the concentrator from being overdrawn when
using with the SureFlow unit.
1.) Supply power to the oxygen concentrator, and set the I/O switch to the I
position to turn it on.
2.) Set the flowmeter on the oxygen concentrator to the maximum capacity of
the unit (for example 5 LPM, 8 LPM, or 10 LPM, depending on
concentrator model).
3.) Connect the oxygen hosing (supplied) to the outlet of the oxygen
concentrator, as shown in Figure 1.
Figure 1
- MN170-1 D (40 pages)
- MN172-1 (502 pages)
- MN189-1 (12 pages)
- MN123-1 F (45 pages)
- MN131-1 B (556 pages)
- MN166-1 B (40 pages)
- MN167-1 (520 pages)
- MN168-1 (3 pages)
- MN169-1 (40 pages)
- MN092-1 D (13 pages)
- MN104-1 D (40 pages)
- MN112-1 (208 pages)
- MN119-1 D (4 pages)
- MN120-1 (44 pages)
- MN078-1 F (42 pages)
- MN084-1 D (137 pages)
- MN139-1 (144 pages)
- MN164-1 (30 pages)
- MN136-1 H (27 pages)
- MN137-1 D (190 pages)
- MN156-1 B (18 pages)
- MN157-1 B (332 pages)
- MN165-1 G (28 pages)
- MN175-1 (2 pages)
- 15062781 (16 pages)
- 4111-7 (95 pages)
- 20626704 (24 pages)
- 20626704-C1 (154 pages)
- ML-CONC0056 (2 pages)
- MN028-1 (1 page)
- MN171-1 (23 pages)
- MN083-1 (1 page)
- MN126-1 C (69 pages)
- MN179-1 B (62 pages)
- MN110-1 (30 pages)
- MN105-1 C (52 pages)
- MN118-1 B (40 pages)
- MN134-1 B (40 pages)
- MN151-1 (1 page)
- MN188-1 (1 page)
- MN188-2 (2 pages)
- MN155-1 (16 pages)
- MN138-1 H (46 pages)
- MN178-1 (1 page)