10
Units Equipped with Barb Fitting Oxygen Outlets
When connecting oxygen concentrators with a barb fitting as the oxygen outlet,
such as AirSep’s VisionAire Oxygen Concentrator, the oxygen hosing will not be
needed.
Instead, securely connect standard oxygen tubing to the barb fitting on the
oxygen concentrator and then using an oxygen adapter (AirSep part number
F0025-1) on the inlet of the SureFlow flow station, connect the other end of the
tubing. (See Figure 5.)
Figure 5
AirSep VisionAire oxygen concentrator models require a minimum flowrate of 1
LPM to achieve high concentration oxygen. When SureFlow is used with the
VisionAire ensure the combined flowrate is at least 1 LPM. If combined patient use
less than 1 LPM, it will be necessary to vent some of the oxygen to atmosphere
through an unused oxygen outlet on the SureFlow in order to achieve a total
flowrate of at least 1 LPM.
- MN170-1 D (40 pages)
- MN172-1 (502 pages)
- MN189-1 (12 pages)
- MN123-1 F (45 pages)
- MN131-1 B (556 pages)
- MN166-1 B (40 pages)
- MN167-1 (520 pages)
- MN168-1 (3 pages)
- MN169-1 (40 pages)
- MN092-1 D (13 pages)
- MN104-1 D (40 pages)
- MN112-1 (208 pages)
- MN119-1 D (4 pages)
- MN120-1 (44 pages)
- MN078-1 F (42 pages)
- MN084-1 D (137 pages)
- MN139-1 (144 pages)
- MN164-1 (30 pages)
- MN136-1 H (27 pages)
- MN137-1 D (190 pages)
- MN156-1 B (18 pages)
- MN157-1 B (332 pages)
- MN165-1 G (28 pages)
- MN175-1 (2 pages)
- 15062781 (16 pages)
- 4111-7 (95 pages)
- 20626704 (24 pages)
- 20626704-C1 (154 pages)
- ML-CONC0056 (2 pages)
- MN028-1 (1 page)
- MN171-1 (23 pages)
- MN083-1 (1 page)
- MN126-1 C (69 pages)
- MN179-1 B (62 pages)
- MN110-1 (30 pages)
- MN105-1 C (52 pages)
- MN118-1 B (40 pages)
- MN134-1 B (40 pages)
- MN151-1 (1 page)
- MN188-1 (1 page)
- MN188-2 (2 pages)
- MN155-1 (16 pages)
- MN138-1 H (46 pages)
- MN178-1 (1 page)