Product specifications – AirSep MN078-1 F User Manual
Page 26

MN078-1 Rev F
8/12
20
Product Specifications
Oxygen Concentration: *
1-3 lpm: 95.5% - 92.0%
4 lpm: 92% ± 3%
5 lpm: 90% ± 3%
Dimensions:
28.5 in. high x 15.7 in. wide x 14.5 in. deep
(72.4 cm high x 40 cm wide x 36.8 cm deep)
Weight:
54 lb (24.5 kg)
Electrical:
120 VAC, 60 Hz, 4.0 amps
350 watts ± 5 lpm model
Two-prong polarized plug
Double insulated cabinet
CSA-Approved to IEC 601-1
Export Models:
220-240 VAC, 50 Hz, 2.0 amps
220 VAC, 60 Hz, 2.0 amps
Double-insulated cabinet
Alarms:
Battery test
Power failure
High and low pressure
Low oxygen concentration (with Oxygen M onitor option)
Storage Temperature:
-4F to 140F (-20C to 60C)
* Based on an atmospheric pressure of 14.7 psi (101 kPa) at 70F (21C)
* Operating unit outside of operational temperature range can affect performance.
- MN170-1 D (40 pages)
- MN172-1 (502 pages)
- MN189-1 (12 pages)
- MN123-1 F (45 pages)
- MN131-1 B (556 pages)
- MN166-1 B (40 pages)
- MN167-1 (520 pages)
- MN168-1 (3 pages)
- MN169-1 (40 pages)
- MN092-1 D (13 pages)
- MN104-1 D (40 pages)
- MN112-1 (208 pages)
- MN119-1 D (4 pages)
- MN120-1 (44 pages)
- MN084-1 D (137 pages)
- MN139-1 (144 pages)
- MN164-1 (30 pages)
- MN152-1 (18 pages)
- MN136-1 H (27 pages)
- MN137-1 D (190 pages)
- MN156-1 B (18 pages)
- MN157-1 B (332 pages)
- MN165-1 G (28 pages)
- MN175-1 (2 pages)
- 15062781 (16 pages)
- 4111-7 (95 pages)
- 20626704 (24 pages)
- 20626704-C1 (154 pages)
- ML-CONC0056 (2 pages)
- MN028-1 (1 page)
- MN171-1 (23 pages)
- MN083-1 (1 page)
- MN126-1 C (69 pages)
- MN179-1 B (62 pages)
- MN110-1 (30 pages)
- MN105-1 C (52 pages)
- MN118-1 B (40 pages)
- MN134-1 B (40 pages)
- MN151-1 (1 page)
- MN188-1 (1 page)
- MN188-2 (2 pages)
- MN155-1 (16 pages)
- MN138-1 H (46 pages)
- MN178-1 (1 page)